Biofilm in the dental unit: what hydrogen peroxide achieves, where it reaches its limits and what chlorine dioxide does differently
In most dental units, the process water is disinfected with hydrogen peroxide [3], often with added silver ions. This keeps the colony count in the water low as long as it is dosed regularly. According to the available studies, it does not completely remove an established biofilm on the tubing wall. Chlorine dioxide acts via a different chemical pathway. This article puts both active substances into context, citing sources, and describes the purpose for which we offer biostream® ZERO, containing the active substance chlorine dioxide, and where its limits lie.
Technical article by Dipl.-Des. (FH) Hans-Joachim Nolte, CEO of biostream® GmbH. As of:
Technical article as PDF (German, 225 KB)Information sheet for practices, 2 pages (German, 145 KB)
How does the Drinking Water Ordinance classify the two substances?
Which substances may be added when drinking water is treated is set by the German Environment Agency (UBA) in the list under Section 20 of the Drinking Water Ordinance (Trinkwasserverordnung, TrinkwV). The current version is dated October 2025 [1]. Part I c of the list covers the substances „die zur Desinfektion des Wassers eingesetzt werden“ (that are used to disinfect water). Five substances are listed there: calcium hypochlorite, chlorine, chlorine dioxide, sodium hypochlorite and ozone.
Hydrogen peroxide is not listed as a disinfectant in this list. It appears in Part I a under serial number 60, with the purpose “oxidation”. A maximum of 17 mg/l may be added, and a maximum of 0.1 mg/l may remain in the water once treatment is complete [1]. This was already the case in the version of January 2023 [2].
Chlorine dioxide appears in Part I c under number 3 with the purpose “disinfection”. A maximum of 0.4 mg/l may be added, and after treatment the water must contain between 0.05 and 0.2 mg/l. As reaction products to be observed, the list names chlorite and chlorate, each with its own maximum value [1].
| Feature | Hydrogen peroxide | Chlorine dioxide |
|---|---|---|
| Location in the list | Part I a, No. 60 | Part I c, No. 3 |
| Purpose | Oxidation | Disinfection |
| Maximum addition | 17 mg/l | 0.4 mg/l |
| After treatment | maximum 0.1 mg/l | 0.05 to 0.2 mg/l |
| Reaction products to be observed | none named | chlorite, chlorate |
For a complete picture, two points must be added. The list concerns drinking water, and substances are included on application. The fact that a substance is not listed among the disinfectants does not mean that it has no disinfecting effect. Hydrogen peroxide is approved in the EU as an active substance for disinfectants [4], and the AWMF guideline (AWMF: Association of the Scientific Medical Societies in Germany) confirms that it acts „gegenüber prokaryontischen Kleinlebewesen desinfizierend“ (as a disinfectant against prokaryotic microorganisms) [3].
Does the Drinking Water Ordinance apply to the water in the dental unit?
Up to the unit, yes; inside the unit, no. The dental unit must be supplied with drinking water. For a connected appliance, the ordinance names the appliance’s backflow prevention device as the point at which its requirements must be met (Section 10 No. 2 TrinkwV) [26]. In a dental unit this is, for example, an air gap (free outlet) to DIN EN 1717 that separates the unit’s water from the mains supply [3].
The ordinance does not apply to the water beyond it. Section 1 (2) No. 4 TrinkwV excludes water located „in Fließrichtung hinter“ (downstream of) the backflow prevention device of a connected appliance [26]. The official explanatory memorandum expressly names „Zahnarztstühle“ (dental chairs) as an example [27]. Hydrogen peroxide may therefore be used in the unit.
Is “Betriebswasser” (process water) a legal term?
Not for the dental unit. The AWMF guideline calls the water in the unit process water (German: Betriebswasser) [3], and the Robert Koch Institute calls it „sogenanntes Betriebswasser“ (so-called process water) [30]. But no law has defined this term for the dental unit. The word appears neither in the Drinking Water Ordinance nor in the Infection Protection Act (Infektionsschutzgesetz, IfSG) nor in the Medical Devices Implementation Act (Medizinprodukterecht-Durchführungsgesetz, MPDG) [26, 28, 29].
A full-text search of all federal law on 25 November 2026 produced twelve hits in other regulations: in the Water Supply Security Act (Wassersicherstellungsgesetz) and its ordinances for a state of defence [32], in the general conditions for water supply [33] and in the Wastewater Ordinance (Abwasserverordnung) [34]. None of these regulations defines what process water is, and none concerns medical devices. For the dental unit, “Betriebswasser” is therefore a technical term. Legally, the only thing that counts is that the Drinking Water Ordinance does not apply downstream of the backflow prevention device and that infection protection law and medical device law apply there.
What applies to the water in the unit?
It is not unregulated. Dental practices must take the „nach dem Stand der medizinischen Wissenschaft erforderlichen Maßnahmen“ (measures required according to the state of medical science) to prevent infections caused by treatment (Section 23 (3) IfSG) [28]. Medical devices may not be operated if they have defects that can endanger patients (Section 11 MPDG) [29], and may only be operated in accordance with the generally recognised rules of technology (Section 4 (1) of the Medical Devices Operator Ordinance, MPBetreibV) [25]. The Robert Koch Institute writes that for water that leaves the unit and comes into contact with the mucous membranes, requirements apply that „an den Vorgaben der Trinkwasserverordnung orientieren“ (are based on the specifications of the Drinking Water Ordinance) [30]. In 2022 the Administrative Court of Minden confirmed that an authority may, on this basis, order water tests at every treatment unit [31].
The dental unit is therefore subject to its own standards, and they are modelled on those for drinking water. The AWMF guideline gives 100 CFU/ml as the warning limit and 500 CFU/ml as the action limit, and Legionella pneumophila should be below 1 CFU/ml [3]. The same guideline advises against silver ions: „Silberionen sollten aber dem Wasser der Dentaleinheit nicht beigegeben werden“ (silver ions should not be added to the water of the dental unit) [3].
Does hydrogen peroxide work against biofilm?
Yes, but to a limited extent. It lowers the colony count in the water when it is dosed continuously and used regularly as a shock treatment. It does not completely remove an established biofilm.
In a laboratory model with a four-week-old mixed biofilm, a daily dose of 0.02% plus a weekly shock with 0.25% kept the colony count below the guide value. The weekly shock alone worked only until the third week, after which the biofilm recovered. A single shock did not completely remove the biofilm [5].
On tubing with biofilm 10 to 14 years old, 3% and 7% hydrogen peroxide detached the deposit within 24 hours except for a small residue. The authors write: “7% H2O2 (very high concentration) did not eradicate biofilms in the waterline, even when in contact with the line surface for 24 hours” [6]. They rated hydrogen peroxide as useful for biofilm control when used for regular cleaning at 2% and for continuous dosing at 0.05% [6]. In a laboratory test at the University of Duisburg-Essen, 150 mg/l (0.015%) over 24 hours showed neither a killing nor a detaching effect on biofilm, not even with 150 µg/l silver ions [7].
In practice this shows up as recolonisation. In a 24-month study in a hospital dental clinic and three dental practices, shock disinfection with 3% hydrogen peroxide had “a limited effect, with a recolonization period of about 4 weeks” [8]. In a study from Dublin, two peroxide-based products kept the colony count below 200 CFU/ml with weekly use and markedly reduced the biofilm. However, it grew back quickly as soon as the weekly disinfection ended [9].
A clinical pilot study from Beijing (eight treatment units, six weeks) first removed the biofilm with a commercially available cleaning kit. After that, hydrogen peroxide with a shock every two weeks met the limits throughout, but with a shock every three weeks it did so in only 78% of the samples. The authors note as a limitation that their results cannot be transferred to units with established biofilm [10].
Why does hydrogen peroxide make little headway in biofilm?
One important reason is an enzyme. Many bacteria produce catalase, which splits hydrogen peroxide into water and oxygen. In biofilm the bacteria are densely packed, and the outer layers break down the hydrogen peroxide before it gets deeper. Measurements with microelectrodes showed this directly: a flow containing hydrogen peroxide “was unable to penetrate or kill wild-type biofilms” but did penetrate biofilms of bacteria without catalase [11]. The authors conclude: “Catalase protects aggregated bacteria by preventing full penetration of hydrogen peroxide into the biofilm.” [11] An earlier study by the same group found that around 90% of the biofilm cells remained viable in such a flow [12].
Catalase is not the only protection. Biofilm cells are also less susceptible when the agent does reach them, and this applies equally to chlorine-containing agents [13]. For bacteria from dental units grown as biofilm in the laboratory, the concentration that removed the biofilm was 256 times the concentration that killed free-floating cells [14].
According to the available data, silver ions do little to change this. In the Duisburg test, adding silver did not improve the effect on the biofilm [7].
What does chlorine dioxide do differently?
Chlorine dioxide is a different substance with different chemistry. Catalase splits hydrogen peroxide; chlorine dioxide is not among the substances this enzyme breaks down. This one protective mechanism of the biofilm therefore does not apply to chlorine dioxide. The other protective mechanisms apply here too [13], and chlorine dioxide is also consumed in the biofilm [19].
Studies with chlorine dioxide products from other manufacturers and at different concentrations show an effect on biofilm. For biostream® ZERO itself, there is no proof of an effect on biofilm.
- In a dissertation at the University of Wuppertal (Bergische Universität Wuppertal), 60 mg/l chlorine dioxide reduced the living cells in young biofilm of one bacterial species by two orders of magnitude in 10 minutes and by more than four in 20 minutes. In comparative tests with up to 60 mg/l and a contact time of 10 minutes, hydrogen peroxide showed „keine ablösenden Eigenschaften“ (no detaching properties) [15].
- A pilot study in a dental practice compared 5 ppm chlorine dioxide with 0.24% hydrogen peroxide. For the colony count, chlorine dioxide was “no less than” hydrogen peroxide, and in removing biofilm, as seen under the microscope, it was stronger [16].
- A clinical hygiene study from 2026 found that chlorine dioxide “was effective in biofilm removal”. A continuous low dose (1.2 mg/l) lowered the colony count lastingly, whereas a single high shock (22.7 mg/l) lowered it only temporarily. The authors therefore recommend continuous use [17].
- In five dental units over three months, two continuous dosing methods, one with hydrogen peroxide and silver and one with chlorine dioxide, performed better than disinfection between patients. According to the authors, it was not possible to tell whether this was due to the method or to the agent [18].
Chlorine dioxide also has limits. In biofilm from dairy process pipes, at 25 mg/l it was used up at a depth of 100 micrometres [19]. The BMBF joint project „Biofilme in der Trinkwasser-Installation“ (Biofilms in drinking water installations) notes that bacteria in compact biofilm can „eine Kombination aus Reinigung und chemischer Anlagen-Desinfektion selbst mit hohen Chlordioxid-Konzentrationen überleben“ (survive a combination of cleaning and chemical system disinfection even with high chlorine dioxide concentrations) [20]. A single short application is therefore not enough for old deposits.
Why do we disinfect twice on first use?
Where contamination has been detected, hygienic quality is unclear or the system has stood unused for a longer period, our dosing instructions specify a contact time of at least 8 hours, followed by complete emptying and a second disinfection with 1 hour of contact time [21]. We assume that the first pass attacks and loosens the deposit and the second reaches what lay beneath. This is our experience; we have no proof of our own for it on biofilm in dental units.
The studies above are consistent with this procedure. In the Wuppertal work, the biofilm grew back to its starting value within 48 hours after 10 minutes of chlorine dioxide, while after 20 minutes it stayed more than two orders of magnitude below it [15]. A longer contact time therefore had a more lasting effect there. None of the studies investigated whether a second pass provides additional benefit.
The AWMF guideline urges caution when removing biofilm. It points out „dass hierzu wissenschaftliche Daten fehlen“ (that scientific data are lacking on this) and that detaching biofilm „zu massiv erhöhten Koloniezahlen kommen kann, die den Betrieb der Dentaleinheit unmöglich machen können“ (can lead to massively increased colony counts that can make operating the dental unit impossible) [3]. It recommends carrying out a biofilm reduction only after careful consideration and on the basis of the manufacturer’s experience with the respective type of device [3]. You should therefore agree the initial disinfection with the manufacturer of your unit, rinse thoroughly afterwards, and have a water sample tested before you use the unit for patients again.
biostream® ZERO at a glance
biostream® ZERO is a two-component chlorine dioxide solution that you combine in the original container before first use. It is a biocidal product and is notified to the German Federal Institute for Occupational Safety and Health (BAuA), registration number N-23862. Chlorine dioxide is still under review in the EU as an active substance for disinfectants. Until a decision is made, notified products containing this active substance may be sold and used [22].
The efficacy against germs in water was tested by the Institute of Food Science and Human Nutrition at Leibniz University Hannover in 1,620 individual tests, commissioned by us. At a contact time of 5 minutes, biostream® ZERO achieved a 99.9% germ reduction (three orders of magnitude) against germs found in drinking water, including Legionella pneumophila and Pseudomonas aeruginosa [23]. This test concerned germs in the water, not biofilm.
biostream® ZERO is intended for disinfecting the process water and the waterlines of treatment units: once at the start for the whole system, then continuously via the disinfection unit or a bottle-care system [24]. Quantities and contact times are given in the dosing instructions and on the label [21]. After a shock disinfection, containers and lines are emptied completely and refilled with drinking water before patients are treated [21].
What should a practice clarify before switching?
No manufacturer of treatment units has given clearance for biostream® ZERO [24]. Under the Medical Devices Operator Ordinance, medical devices may only be operated in accordance with their intended purpose, and connected devices only if they are suitable in that combination [25]. Please therefore check the instructions for use of your unit, and clarify with the manufacturer before use whether you may use a different agent, also with regard to the warranty.
The AWMF guideline requires material compatibility to be „kritisch zu überprüfen“ (critically checked) for chlorine dioxide [3]. Keep to the dosing quantities and contact times in the instructions, and have the water of your unit tested microbiologically after the changeover. The guideline’s guide values are the yardstick for this [3].
We answer questions about use in your practice at info@biostream.de or by telephone on +49 5138 708 78 38.
Use biocides safely. Always read the label and product information before use.
Sources
- Umweltbundesamt: 1. Änderung der Liste zulässiger Aufbereitungsstoffe und Desinfektionsverfahren gemäß § 20 der Trinkwasserverordnung (Stand: Oktober 2025), vom 23.09.2025. BAnz AT 10.10.2025 B7. https://www.bundesanzeiger.de/pub/
publication/ uesJg7QaDKr5EPuvJMw/ content/ uesJg7QaDKr5EPuvJMw/ BAnz%20AT%2010.10.2025%20B7.pdf - Umweltbundesamt: Bekanntmachung der Liste zulässiger Aufbereitungsstoffe und Desinfektionsverfahren gemäß § 20 der Trinkwasserverordnung (Stand: Januar 2023). BAnz AT 27.01.2023 B12. https://www.umweltbundesamt.de/themen/
wasser/ trinkwasser/ rechtliche-grundlagen-empfehlungen-regelwerk/ aufbereitungsstoffe-desinfektionsverfahren-ss-20 - DGKH, DGZMK: S2k-Leitlinie „Dentaleinheiten, hygienische Anforderungen an das Wasser“, AWMF-Register 075-002, Version 2.1, Stand 28.02.2026. https://register.awmf.org/de/leitlinien/
detail/075-002 - Durchführungsverordnung (EU) 2015/1730 zur Genehmigung von Wasserstoffperoxid als Wirkstoff zur Verwendung in Biozidprodukten. https://eur-lex.europa.eu/eli/
reg_impl/2015/1730/oj - Hoogenkamp MA, Mazurel D, Deutekom-Mulder E, de Soet JJ: The consistent application of hydrogen peroxide controls biofilm growth and removes Vermamoeba vermiformis from multi-kingdom in-vitro dental unit water biofilms. Biofilm 2023;5:100132. https://doi.org/10.1016/j.bioflm.2023.100132
- Lin SM, Svoboda KKH, Giletto A, Seibert J, Puttaiah R: Effects of hydrogen peroxide on dental unit biofilms and treatment water contamination. Eur J Dent 2011;5(1):47–59. https://pmc.ncbi.nlm.nih.gov/
articles/ PMC3019751/ - Schulte S: Wirksamkeit von Wasserstoffperoxid gegenüber Biofilmen. Dissertation, Universität Duisburg-Essen, 2003. https://duepublico2.uni-due.de/servlets/
MCRFileNodeServlet/ duepublico_derivate_00005510/index.html - Tuvo B, Totaro M, Cristina ML et al.: Prevention and Control of Legionella and Pseudomonas spp. Colonization in Dental Units. Pathogens 2020;9(4):305. https://doi.org/10.3390/pathogens9040305
- Tuttlebee CM, O'Donnell MJ, Keane CT et al.: Effective control of dental chair unit waterline biofilm and marked reduction of bacterial contamination of output water using two peroxide-based disinfectants. J Hosp Infect 2002;52:192–205. https://doi.org/10.1053/jhin.2002.1282
- Wen J, Liu W, Shao T et al.: Biweekly versus triweekly hydrogen peroxide shock treatment for controlling dental unit waterline contamination: a single-center pilot randomized clinical study. Front Oral Health 2026;7:1842649. https://doi.org/10.3389/froh.2026.1842649
- Stewart PS, Roe F, Rayner J et al.: Effect of catalase on hydrogen peroxide penetration into Pseudomonas aeruginosa biofilms. Appl Environ Microbiol 2000;66:836–838. https://doi.org/10.1128/AEM.66.2.836-838.2000
- Elkins JG, Hassett DJ, Stewart PS et al.: Protective role of catalase in Pseudomonas aeruginosa biofilm resistance to hydrogen peroxide. Appl Environ Microbiol 1999;65:4594–4600. https://doi.org/10.1128/aem.65.10.4594-4600.1999
- Cochran WL, McFeters GA, Stewart PS: Reduced susceptibility of thin Pseudomonas aeruginosa biofilms to hydrogen peroxide and monochloramine. J Appl Microbiol 2000;88:22–30. https://doi.org/10.1046/j.1365-2672.2000.00825.x
- Yoon HY, Lee SY: Susceptibility of bacteria isolated from dental unit waterlines to disinfecting chemical agents. J Gen Appl Microbiol 2019;64:269–275. https://doi.org/10.2323/jgam.2018.02.001
- Czirr JK: Quantifizierung und Reduzierung von Biofilmen verschiedener Trinkwasserisolate. Dissertation, Bergische Universität Wuppertal, 2017. https://elekpub.bib.uni-wuppertal.de/ubwhsmig/
download/ pdf/3555227 - Zhang X, Sha J, Huang Z et al.: A pilot study comparing the disinfecting effects of commercialized stable ClO2 solution (free of activation) with conventional H2O2 on dental unit waterlines in the dental practice setting. Drug Discov Ther 2023;17:357–362. https://doi.org/10.5582/ddt.2023.01077
- Winkler J, Herzog S, Dahlhaus F et al.: Chlorine dioxide flushing protocols for microbial reduction in dental chair units. PLoS One 2026;21:e0342347. https://doi.org/10.1371/journal.pone.0342347
- Dallolio L, Scuderi A, Rini MS et al.: Effect of different disinfection protocols on microbial and biofilm contamination of dental unit waterlines in community dental practices. Int J Environ Res Public Health 2014;11:2064–2076. https://doi.org/10.3390/ijerph110202064
- Jang A, Szabo J, Hosni AA et al.: Measurement of chlorine dioxide penetration in dairy process pipe biofilms during disinfection. Appl Microbiol Biotechnol 2006;72:368–376. https://doi.org/10.1007/s00253-005-0274-5
- Erkenntnisse aus dem BMBF-Verbundprojekt „Biofilme in der Trinkwasser-Installation“ (Koordination Prof. Dr. Hans-Curt Flemming, Projektlaufzeit 2006–2010, Partner u. a. Institut für Hygiene und Öffentliche Gesundheit der Universität Bonn), Thesenpapier Version 2.1. https://comprex.de/wp-content/
uploads/2023/03/Thesenpapier_2_0.pdf - biostream GmbH: Dosieranleitung Chlordioxid. https://www.biostream.de/dosierungsanleitung-chlordioxid/
- REACH-CLP-Biozid Helpdesk der Bundesbehörden: Übergangsregelungen für Biozidprodukte. https://www.reach-clp-biozid-helpdesk.de/de/Biozide/
Vorschriften/ Uebergangsregelungen - biostream GmbH: Wirksamkeitsnachweis, Institut für Lebensmittelwissenschaft und Ökotrophologie der Leibniz Universität Hannover. https://www.biostream.de/wirksamkeitsnachweis/
- biostream GmbH: Desinfektion von Behandlungseinheiten bei Zahnärzten. https://www.biostream.de/biostream-zero/
desinfektion-von-behandlungseinheiten-bei-zahnaerzten/ - Medizinprodukte-Betreiberverordnung, § 4 Abs. 1 und 4. https://www.gesetze-im-internet.de/mpbetreibv_2025/__4.html
- Trinkwasserverordnung vom 20.06.2023, § 1 Abs. 2 Nr. 4 und § 10 Nr. 2. https://www.gesetze-im-internet.de/trinkwv_2023/
- Bundesrat, Drucksache 530/10 vom 02.09.2010, Erste Verordnung zur Änderung der Trinkwasserverordnung, Begründung zu § 2 Abs. 1 Satz 2 Nr. 4, S. 51. https://dserver.bundestag.de/brd/2010/0530-10.pdf
- Infektionsschutzgesetz, § 23 Abs. 3. https://www.gesetze-im-internet.de/ifsg/__23.html
- Medizinprodukterecht-Durchführungsgesetz, § 11. https://www.gesetze-im-internet.de/mpdg/__11.html
- Robert Koch-Institut: Infektionshygiene in der Zahnheilkunde, Frage zur Bewertung von Wasser aus Dentaleinheiten. https://www.rki.de/DE/Themen/
Infektionskrankheiten/ Krankenhaushygiene/ Infektionshygiene-A-Z/Z/Zahnheilkunde/ zahnheilkunde-inhalt.html - Verwaltungsgericht Minden, Urteil vom 15.03.2022, 7 K 3145/20. https://nrwe.justiz.nrw.de/ovgs/vg_minden/j2022/7_K_3145_20_Urteil_20220315.html
- Wassersicherstellungsgesetz, §§ 1 bis 3; Erste Wassersicherstellungsverordnung, §§ 4 und 5. https://www.gesetze-im-internet.de/wassv_1/__5.html
- Verordnung über Allgemeine Bedingungen für die Versorgung mit Wasser, § 4 Abs. 3. https://www.gesetze-im-internet.de/avbwasserv/__4.html
- Abwasserverordnung, Anhänge 31, 35 und 40. https://www.gesetze-im-internet.de/abwv/
Related topics: Disinfecting the waterlines of the treatment unit, Fighting biofilms, Chlorine dioxide