Feedback from dental practices
Dentists regularly contact biostream® GmbH when findings such as Legionella have been detected in treatment units, asking how to fight them effectively. The treatment unit has to be disinfected as quickly as possible to avoid an interruption to operations. Unfortunately, support from dental depots often does not bring the expected solution, because they often lack the understanding and knowledge of the subject "disinfection of drinking water in treatment units". On closer inspection, hydrogen peroxide solutions are often recommended, which under the current Drinking Water Ordinance are not named as a disinfectant at all.
Our work with dentists has shown that there are many problems in the disinfection of treatment units. Below we report on a case that may sound familiar to you too. (We add to these examples regularly.)
Case 1: Legionella despite correct application, caused by dead legs
A dentist from Göttingen, who had already been dealing with the subject of "disinfecting treatment units" for years, had a problem with Legionella after switching to the bottle-care system. After using biostream® ZERO according to our dosing recommendation, no more Legionella could be detected. The dentist was very satisfied and let us know so shortly afterwards. After a week, however, recontamination of the treatment unit was found, and the dentist's feedback was correspondingly negative. The efficacy of biostream® ZERO for disinfecting treatment units at dental practices was called into question.
Because fresh contamination with Legionella within a week of application is extremely unusual and had not occurred before, our assumption was that there had to be so-called dead legs directly in the treatment unit. Dead legs are sections of tubing with no fitting connected at the end.
The dentist then contacted his dental depot and asked whether there were dead legs in his dental unit. The dental laboratory's answer was: "No!" The dentist then requested a thorough on-site inspection of his unit and agreed to bear the cost himself if NO dead legs were found. If dead legs were confirmed, the dental depot was to bear the cost. The dental depot agreed, and a complete inspection took place shortly afterwards. It turned out that the dental unit contained 2 × 30 cm of water tubing that had been installed for operating a second treatment unit. However, since the dentist did not need a second treatment unit, it had never been fitted. The 2 × 30 cm water hoses were nevertheless still there.
And it was exactly these hoses that had caused disinfection to fail: effective disinfection of treatment units can only work if ALL the drinking-water-carrying lines can be wetted with disinfectant.
Disinfection is not possible in dead legs! Why? In a dead leg, a disinfectant cannot reach the water column. It is a dead end: there is no flow that can carry a disinfectant into this dead leg. For this reason, biostream® ZERO was also unable to fight the Legionella in the dead legs in the case described.
In line with the agreement, the dental depot naturally bore the cost of the inspection. After removing the 2 × 30 cm dead legs, no more Legionella could be detected.